<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>30012</ErrorNumber>
<ErrorFamily>Test Pattern Generator</ErrorFamily>
<ErrorMessage>
TPG: timed out waiting for stripe to complete
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The test pattern generator (TPG) on the DPE is not functioning properly. The TPG
is used at the beginning and end of every image to setup and shutdown the data
buffers.
</text></DescriptionText>
<DescriptionText><text>
If this error occurs at the beginning of an image, the DPE should be re-seated
and all power supply levels (including the -5.2V and +4.6V ECL supplies) should
be checked using the test points on board. If the problem does not go away, the
DPE should be replaced.
</text></DescriptionText>
<DescriptionText><text>
If this error occurs at the end of an image, the DPE should be checked. However,
the error may be due to a firmware bug in the image shutdown routine, and the
responsible firmware programmer should be contacted.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Record error message and contact Creo service
</text></DescriptionText>
</OperatorAction>
<Cause>
Out of spec. power supply levels
</Cause>
<Cause>
Faulty DPE
</Cause>
<Cause>
If at the end of an image, possibly due to a firmware bug.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:34:16 2008 PST
</Updated>
<Links>
</Links>
</Error>